Study of focused ultrasound tissue damage using MRI and histology.
Authors: Chen L, Bouley D, Yuh E, D'Arceuil H, Butts K
This paper reports on an experimental study of in vivo tissue damage in the rabbit brain with focused ultrasound (FUS) using magnetic resonance imaging (MRI) and histopathological analysis. Ten ultrasonic lesions (tissue damage) were created in five rabbits using a focused ultrasound beam of 1.5 MHz, electrical power input to the transducer of 70-85 W, and an exposure duration of 15-20 seconds. T1- and T2-weighted fast spin-echo (FSE) and Fluid attenuated inversion recovery (FLAIR) sequences were used to detect the ultrasonic lesions after treatment. Imaging was performed for 4-8 hours after treatment, after which the animals were immediately sacrificed. Ultrasonic lesion diameter was measured on MRI and histological sections after correction for tissue shrinkage during the histological processing. The T1-weighted images showed lesions poorly, whereas both T2-weighted and FLAIR images showed lesions clearly. The lesion diameters on both T2 and FLAIR imaging correlated well with measurements from histology. The time delay before lesions appeared on T2-weighted imaging was 15 minutes to 1 hour, depending on the exposure location in the brain. J. Magn. Reson. Imaging 1999;10:146-153.
Introduction
Purpose
Thermal ablation
Study Objective
To evaluate tissue damage caused by focused ultrasound using MRI and histological analysis.
Animal model / Human subject
rabbit
Disease model
Healthy
MRI or image guidance method
MRI (T1-weighted, T2-weighted FSE, FLAIR)
Outcomes and Safety
Summary of Outcomes
FUS-induced brain lesions successfully characterized using MRI and histology. T2-weighted and FLAIR imaging showed good agreement with histological lesion measurements and outperformed T1-weighted imaging.
Duration of biological effect
4-8 h
Safety-related matter
FUS created localized tissue damage as intended for lesion formation. No additional adverse effects or safety findings were reported in the abstract.
Brain Region
Ultrasound Parameters
FUS Frequency
1.5 Mhz
Duration of a single FUS session
15-20 seconds
Focal Characteristics
Focal depth: None; Focal length: None; Aperture size: None
Treatment frequency
single session