Focused ultrasound combined with radiotherapy for malignant brain tumor: a preclinical and clinical study.
Authors: Chen KT, Huang CY, Pai PC, Yang WC, Tseng CK, Tsai HC, Li JC, Chuang CC, Hsu PW, Lee CC, Toh CH, Liu HL, Wei KC
Blood-brain barrier (BBB) remains to be the major obstacle to conquer in treating patients with malignant brain tumors. Radiation therapy (RT), despite being the mainstay adjuvant modality regardless of BBB, the effect of radiation induced cell death is hindered by the hypoxic microenvironment. Focused ultrasound (FUS) combined with systemic microbubbles has been shown not only to open BBB but also potentially increased regional perfusion. However, no clinical study has investigated the combination of RT with FUS-BBB opening (RT-FUS). We aimed to provide preclinical evidence of RT-FUS combination in GBM animal model, and to report an interim analysis of an ongoing single arm, prospective, pilot study (NCT01628406) of combining RT-FUS for recurrent malignant high grade glioma patients, of whom re-RT was considered for disease control. In both preclinical and clinical studies, FUS-BBB opening was conducted within 2 h before RT. Treatment responses were evaluated by objective response rate (ORR) using magnetic resonance imaging, progression free survival, and overall survival, and adverse events (AE) in clinical study. Survival analysis was performed in preclinical study and descriptive analysis was performed in clinical study. In mouse GBM model, the survival analysis showed RT-FUS (2 Gy) group was significantly longer than RT (2 Gy) group and control, but not RT (5 Gy) group. In the pilot clinical trial, an interim analysis of six recurrent malignant high grade glioma patients underwent a total of 24 RT-FUS treatments was presented. Three patients had rapid disease progression at a mean of 33 days after RT-FUS, while another three patients had at least stable disease (mean 323 days) after RT-FUS with or without salvage chemotherapy or target therapy. One patient had partial response after RT-FUS, making the ORR of 16.7%. There was no FUS-related AEs, but one (16.7%) re-RT-related grade three radiation necrosis. Reirradiation is becoming an option after disease recurrence for both primary and secondary malignant brain tumors since systemic therapy significantly prolongs survival in cancer patients. The mechanism behind the synergistic effect of RT-FUS in preclinical model needs further study. The clinical evidence from the interim analysis of an ongoing clinical trial (NCT01628406) showed a combination of RT-FUS was safe (no FUS-related adverse effect). A comprehensive analysis of radiation dosimetry and FUS energy distribution is expected after completing the final recruitment.
Introduction
Purpose
Drug delivery with BBB opening
Study Objective
To investigate the presence and impact of standing waves during microbubble-mediated focused ultrasound blood–brain barrier disruption in rats and to develop pulse and transducer methods to eliminate them.
Animal model / Human subject
Rat (Rattus norvegicus); strain: not specified; age: not specified; sex: not specified
Disease model
Healthy
Outcomes and Safety
Summary of Outcomes
Microbubble-mediated focused ultrasound successfully disrupted the blood–brain barrier in vivo in rats and this effect did not depend on the presence of standing waves; standing waves were common ex vivo and could not be eliminated with swept-frequency excitation. Standing waves were successfully eliminated using a wideband composite sharply-focused transducer and a reduced duty cycle (modified pulse parameters), whereas swept frequencies were ineffective.
Safety-related matter
No adverse effects were reported. The authors warn that standing waves cause high variability and make accurate in situ pressure estimation difficult, so minimizing standing waves is recommended to ensure clinical translatability and safety.
Brain Region
Ultrasound Parameters
Ultrasound instrument
Wideband composite transducer
FUS Frequency
1.503 MHz
FUS Pressure
0.58 Mpa; 0.71 Mpa
FUS Mode
pulsed
Pulse duration
10 ms
Duration of a single FUS session
2 min
Focal Characteristics
Focal depth: None; Focal length: 80 mm; Aperture size: 100 mm
Treatment frequency
Multiple