Permeability dependence study of the focused ultrasound-induced blood-brain barrier opening at distinct pressures and microbubble diameters using DCE-MRI.

Authors: Vlachos F, Tung YS, Konofagou E

Blood-brain barrier opening using focused ultrasound and microbubbles has been experimentally established as a noninvasive and localized brain drug delivery technique. In this study, the permeability of the opening is assessed in the murine hippocampus after the application of focused ultrasound at three different acoustic pressures and microbubble sizes. Using dynamic contrast-enhanced MRI, the transfer rates were estimated, yielding permeability maps and quantitative K(trans) values for a predefined region of interest. The volume of blood-brain barrier opening according to the K(trans) maps was proportional to both the pressure and the microbubble diameter. A K(trans) plateau of ∼0.05 min(-1) was reached at higher pressures (0.45 and 0.60 MPa) for the larger sized bubbles (4-5 and 6-8 μm), which was on the same order as the K(trans) of the epicranial muscle (no barrier). Smaller bubbles (1-2 μm) yielded significantly lower permeability values. A small percentage (7.5%) of mice showed signs of damage under histological examination, but no correlation with permeability was established. The assessment of the blood-brain barrier permeability properties and their dependence on both the pressure and the microbubble diameter suggests that K(trans) maps may constitute an in vivo tool for the quantification of the efficacy of the focused ultrasound-induced blood-brain barrier opening.

Introduction

Purpose Drug delivery with BBB opening
Study Objective To assess and quantify blood–brain barrier permeability in the murine hippocampus following focused ultrasound with varying acoustic pressures and microbubble sizes using DCE-MRI-derived Ktrans maps.
Animal model / Human subject Mouse (Mus musculus); strain: None; age: None; sex: None
Disease model Healthy
Targeted brain region(s) Hippocampus
Cargo name and characteristics DCE‑MRI contrast agent (small‑molecule gadolinium‑based contrast agent)

Outcomes and Safety

Summary of Outcomes Focused ultrasound with microbubbles produced localized BBB opening in the murine hippocampus; permeability (Ktrans) and opening volume increased with acoustic pressure and microbubble diameter, reaching a plateau Ktrans ≈ 0.05 min^-1 at higher pressures. Effective parameters were higher pressures (0.45 and 0.60 MPa) with larger microbubbles (4–5 µm and 6–8 µm); smaller bubbles (1–2 µm) yielded significantly lower permeability and 7.5% of mice showed histological damage with no correlation to permeability.
Safety-related matter A small percentage (7.5%) of mice showed signs of damage on histological examination; no correlation between the observed damage and BBB permeability (Ktrans) was established.