MR-guided adaptive focusing of therapeutic ultrasound beams in the human head.

Authors: Marsac L, Chauvet D, Larrat B, Pernot M, Robert B, Fink M, Boch AL, Aubry JF, Tanter M

This study aims to demonstrate, using human cadavers the feasibility of energy-based adaptive focusing of ultrasonic waves using magnetic resonance acoustic radiation force imaging (MR-ARFI) in the framework of non-invasive transcranial high intensity focused ultrasound (HIFU) therapy. Energy-based adaptive focusing techniques were recently proposed in order to achieve aberration correction. The authors evaluate this method on a clinical brain HIFU system composed of 512 ultrasonic elements positioned inside a full body 1.5 T clinical magnetic resonance (MR) imaging system. Cadaver heads were mounted onto a clinical Leksell stereotactic frame. The ultrasonic wave intensity at the chosen location was indirectly estimated by the MR system measuring the local tissue displacement induced by the acoustic radiation force of the ultrasound (US) beams. For aberration correction, a set of spatially encoded ultrasonic waves was transmitted from the ultrasonic array and the resulting local displacements were estimated with the MR-ARFI sequence for each emitted beam. A noniterative inversion process was then performed in order to estimate the spatial phase aberrations induced by the cadaver skull. The procedure was first evaluated and optimized in a calf brain using a numerical aberrator mimicking human skull aberrations. The full method was then demonstrated using a fresh human cadaver head. The corrected beam resulting from the direct inversion process was found to focus at the targeted location with an acoustic intensity 2.2 times higher than the conventional non corrected beam. In addition, this corrected beam was found to give an acoustic intensity 1.5 times higher than the focusing pattern obtained with an aberration correction using transcranial acoustic simulation-based on X-ray computed tomography (CT) scans. The proposed technique achieved near optimal focusing in an intact human head for the first time. These findings confirm the strong potential of energy-based adaptive focusing of transcranial ultrasonic beams for clinical applications.

Introduction

Purpose Thermal ablation
Study Objective To demonstrate the feasibility of energy-based adaptive focusing of transcranial ultrasonic waves using MR-ARFI for aberration correction in non-invasive HIFU therapy on human cadaver heads.
Animal model / Human subject Human cadaver head (Homo sapiens): strain None; age None; sex None; Calf brain (calf, Bos taurus): strain None; age: calf (None); sex None
MRI or image guidance method MR-guided using Magnetic Resonance Acoustic Radiation Force Imaging (MR-ARFI) in a 1.5 T MRI, with cadaver heads mounted on a Leksell stereotactic frame

Outcomes and Safety

Summary of Outcomes Energy-based adaptive focusing using MR-ARFI with a direct inversion aberration correction produced near-optimal transcranial HIFU focusing, increasing acoustic intensity 2.2× versus the non-corrected beam and 1.5× versus CT-simulation-based correction (successful approach: MR-ARFI direct inversion; comparators tested: non-corrected and CT-based aberration correction).
Safety-related matter No safety concerns or adverse effects are reported in the paper; the experiments were conducted on human cadaver heads and no clinical safety or adverse-event data are provided.