Pitt Shield

Opening of the blood-brain barrier with an unfocused ultrasound device in rabbits.

Authors: Beccaria K, Canney M, Goldwirt L, Fernandez C, Adam C, Piquet J, Autret G, Clément O, Lafon C, Chapelon JY, Carpentier A

The blood-brain barrier (BBB) is a major impediment to the intracerebral diffusion of drugs used in the treatment of gliomas. Previous studies have demonstrated that pulsed focused ultrasound (US) in conjunction with a microbubble contrast agent can be used to open the BBB. To apply the US-induced opening of the BBB in clinical practice, the authors designed an innovative unfocused US device that can be implanted in the skull and used to transiently and repeatedly open the BBB during a standard chemotherapy protocol. The goal of this preliminary work was to study the opening of the BBB induced by the authors' small unfocused US transducer and to evaluate the effects of the sonications on brain parenchyma. Craniectomy was performed in 16 healthy New Zealand White rabbits; epidural application of a single-element planar ultrasonic transducer operating at 1 MHz was then used with a pulse-repetition frequency of 1 Hz, pulse lengths of 10-35 msec, in situ acoustic pressure levels of 0.3-0.8 MPa, and sonication for 60-120 seconds. SonoVue was intravenously injected during the US applications, and opening of the BBB was determined by detecting extravasation of Evans blue dye (EBD) in brain tissues, quantitative measurement of EBD with UV-visible spectrophotometry, and contrast enhancement after Gd injection in 4.7-T MRI. A histological study was performed to determine adverse effects. An opening of the BBB was observed over a large extent of the US beam in the brain corresponding to in situ pressures of greater than 0.2 MPa. The BBB opening observed was highly significant for both EBD (p < 0.01) and MRI Gd enhancement (p < 0.0001). The BBB opening was associated with minor adverse effects that included perivascular red blood cell extravasations that were less than 150 μm in size and not visible on MR images. Moderate edema was visible on FLAIR sequences and limited to the extent of the sonication field. The results demonstrate that the BBB can be opened in large areas of the brain in rabbits with low-power, pulsed, and unfocused US with limited damage to healthy tissue.

Introduction

Purpose Drug delivery with BBB opening
Study Objective To evaluate whether an unfocused ultrasound device can open the blood–brain barrier in rabbits.
Animal model / Human subject Rabbits; strain not reported; age not reported; sex not reported
Disease model Healthy

Outcomes and Safety

Summary of Outcomes An unfocused ultrasound device was used to open the blood–brain barrier in rabbits; the main biological effect reported is BBB opening in this animal model. Specific ultrasound parameters tested or found successful are not specified in the provided text.
Safety-related matter No safety or adverse effects are mentioned in the provided text.

Brain Region

Visualization unavailable

Ultrasound Parameters

Focal Characteristics Focal depth: None; Focal length: None; Aperture size: None

We are open to feedback. If you see a mistake or have a suggestion, please contact us.

← Back to Search