Pitt Shield

Transcranial high intensity focused ultrasound therapy guided by 7 TESLA MRI in a rat brain tumour model: a feasibility study.

Authors: Dervishi E, Larrat B, Pernot M, Adam C, Marie Y, Fink M, Delattre JY, Boch AL, Tanter M, Aubry JF

Transcranial high intensity focused ultrasound (HIFU) therapy guided by magnetic resonance imaging (MRI) is a promising approach for the treatment of brain tumours. Our objective is to validate a dedicated therapy monitoring system for rodents for transcranial HIFU therapy under MRI guidance in an in vivo brain tumour model. A dedicated MR-compatible ultrasound therapy system and positioning frame was developed. Three MR-compatible prefocused ultrasonic monoelement transducers were designed, operating at 1.5 MHz and 2.5 MHz with different geometries. A full protocol of transcranial HIFU brain therapy under MRI guidance was applied in n = 19 rats without and n = 6 rats with transplanted tumours (RG2). Different heating strategies were tested. After treatment, histological study of the brain was performed in order to confirm thermal lesions. Relying on a larger aperture and a higher frequency, the 2.5 MHz transducer was found to give better results than other ones. This single element transducer optimised the ratio of the temperature elevation at the focus to the one at the skull surface. Using optimised transducer and heating strategies enabled thermal necrosis both in normal and tumour tissues as verified by histology while limiting overheating in the tissues in contact with the skull. In this study, a system for transcranial HIFU therapy guided by MRI was developed and tested in an in vivo rat brain tumour model. The feasibility of this therapy set-up to induce thermal lesions within brain tumours was demonstrated.

Introduction

Purpose Thermal ablation
Study Objective To assess the feasibility of transcranial high-intensity focused ultrasound therapy guided by 7 Tesla MRI in a rat brain tumour model.
Animal model / Human subject Rat (Rattus norvegicus); strain None; age None; sex None
Disease model brain tumour (rat brain tumour model)
MRI or image guidance method 7 Tesla MRI

Outcomes and Safety

Summary of Outcomes This feasibility study indicates that 7T MRI-guided transcranial high-intensity focused ultrasound can be delivered in a rat brain tumor model to produce targeted lesioning/thermal ablation. The provided text does not report specific ultrasound parameter sets or indicate which parameters (if multiple were tested) were found successful.
Safety-related matter No safety or adverse effects are mentioned in the provided text. Only the paper title is present, so safety information is not available.

Brain Region

Visualization unavailable

Ultrasound Parameters

Focal Characteristics Focal depth: None; Focal length: None; Aperture size: None

We are open to feedback. If you see a mistake or have a suggestion, please contact us.

← Back to Search