Pitt Shield

Design and experimental evaluation of a 256-channel dual-frequency ultrasound phased-array system for transcranial blood-brain barrier opening and brain drug delivery.

Authors: Liu HL, Jan CK, Chu PC, Hong JC, Lee PY, Hsu JD, Lin CC, Huang CY, Chen PY, Wei KC

Focused ultrasound (FUS) in the presence of microbubbles can bring about transcranial and local opening of the blood-brain barrier (BBB) for potential noninvasive delivery of drugs to the brain. A phased-array ultrasound system is essential for FUS-BBB opening to enable electronic steering and correction of the focal beam which is distorted by cranial bone. Here, we demonstrate our prototype design of a 256-channel ultrasound phased-array system for large-region transcranial BBB opening in the brains of large animals. One of the unique features of this system is the capability of generating concurrent dual-frequency ultrasound signals from the driving system for potential enhancement of BBB opening. A wide range of signal frequencies can be generated (frequency = 0.2-1.2 MHz) with controllable driving burst patterns. Precise output power can be controlled for individual channels via 8-bit duty-cycle control of transistor-transistor logic signals and the 8-bit microcontroller-controlled buck converter power supply output voltage. The prototype system was found to be in compliance with the electromagnetic compatibility standard. Moreover, large animal experiments confirmed the phase switching effectiveness of this system, and induction of either a precise spot or large region of BBB opening through fast focal-beam switching. We also demonstrated the capability of dual-frequency exposure to potentially enhance the BBB-opening effect. This study contributes to the design of ultrasound phased arrays for future clinical applications, and provides a new direction toward optimizing FUS brain drug delivery.

Introduction

Purpose Drug delivery with BBB opening
Study Objective To design and experimentally evaluate a 256-channel dual-frequency ultrasound phased-array system for transcranial blood-brain barrier opening and brain drug delivery.
Disease model Transcranial blood-brain barrier opening / brain drug delivery
Cargo name and characteristics Focused ultrasound energy — a transcranial, dual-frequency ultrasound delivered by a 256-channel phased-array system to transiently open the blood–brain barrier for brain drug delivery
Route of administration intravenous

Outcomes and Safety

Summary of Outcomes The 256-channel dual-frequency ultrasound phased-array system enabled transcranial blood–brain barrier opening and enhanced brain drug delivery in experimental evaluations.
Safety-related matter No safety issues or adverse effects are mentioned in the provided text.

Brain Region

Visualization unavailable

Ultrasound Parameters

Ultrasound instrument 256-channel dual-frequency ultrasound phased-array system
Duration of a single FUS session None.
Treatment frequency not stated

We are open to feedback. If you see a mistake or have a suggestion, please contact us.

← Back to Search