Real-time monitoring of focused ultrasound blood-brain barrier opening via subharmonic acoustic emission detection: implementation of confocal dual-frequency piezoelectric transducers.
Authors: Tsai CH, Zhang JW, Liao YY, Liu HL
Burst-tone focused ultrasound exposure in the presence of microbubbles has been demonstrated to be effective at inducing temporal and local opening of the blood-brain barrier (BBB), which promises significant clinical potential to deliver therapeutic molecules into the central nervous system (CNS). Traditional contrast-enhanced imaging confirmation after focused ultrasound (FUS) exposure serves as a post-operative indicator of the effectiveness of FUS-BBB opening, however, an indicator that can concurrently report the BBB status and BBB-opening effectiveness is required to provide effective feedback to implement this treatment clinically. In this study, we demonstrate the use of subharmonic acoustic emission detection with implementation on a confocal dual-frequency piezoelectric ceramic structure to perform real-time monitoring of FUS-BBB opening. A confocal dual-frequency (0.55 MHz/1.1 MHz) focused ultrasound transducer was designed. The 1.1 MHz spherically-curved ceramic was employed to deliver FUS exposure to induce BBB-opening, whereas the outer-ring 0.55 MHz ceramic was employed to detect the subharmonic acoustic emissions originating from the target position. In stage-1 experiments, we employed spectral analysis and performed an energy spectrum density (ESD) calculation. An optimized 0.55 MHz ESD level change was shown to effectively discriminate the occurrence of BBB-opening. Wideband acoustic emissions received from 0.55 MHz ceramics were also analyzed to evaluate its correlations with erythrocyte extravasations. In stage-2 real-time monitoring experiments, we applied the predetermined ESD change as a detection threshold in PC-controlled algorithm to predict the FUS exposure intra-operatively. In stage-1 experiment, we showed that subharmonic ESD presents distinguishable dynamics between intact BBB and opened BBB, and therefore a threshold ESD change level (5.5 dB) can be identified for BBB-opening prediction. Using this ESD change threshold detection as a surrogate to on/off control the FUS exposure in stage-2 experiments, we demonstrated both excellent sensitivity (92%) and specificity (92.3%) in discriminating BBB-opening occurrence can be obtained in animal treatments, while concurrently achieving a high positive predicted value (95.8%). Wideband ESD was also highly correlated with the occurrence and level of erythrocyte extravasations (r (2) = 0.81). The proposed system configuration and corresponding analysis based on subharmonic acoustic emissions has the potential to be implemented as a real-time feedback control structure for reliable indication of intact FUS-BBB opening for CNS brain drug delivery.
Introduction
Purpose
Drug delivery with BBB opening
Study Objective
Implement confocal dual-frequency piezoelectric transducers to enable real-time monitoring of focused ultrasound–induced blood–brain barrier opening via detection of subharmonic acoustic emissions.
Animal model / Human subject
45 male Sprague-Dawley rats (250 - 300 g)
Disease model
healthy
MRI or image guidance method
Subharmonic acoustic emission detection with confocal dual-frequency piezoelectric transducers (real-time acoustic-guided targeting)
Cargo name and characteristics
NR
Outcomes and Safety
Summary of Outcomes
Subharmonic acoustic emission signaled FUS induced BBB opening in real time.
Brain Region
Ultrasound Parameters
Ultrasound instrument
A concentric and confocally-arranged dual-frequency FUS transducer
FUS Frequency
1.1 MHz
FUS Pressure
0.44-0.82
FUS Mode
pulsed
Pulse duration
10 ms
Duration of a single FUS session
90 sec
Treatment frequency
multiple sessions