Pitt Shield

Low-Pressure Burst-Mode Focused Ultrasound Wave Reconstruction and Mapping for Blood-Brain Barrier Opening: A Preclinical Examination.

Authors: Xia J, Tsui PH, Liu HL

Burst-mode focused ultrasound (FUS) exposure has been shown to induce transient blood-brain barrier (BBB) opening for potential CNS drug delivery. FUS-BBB opening requires imaging guidance during the intervention, yet current imaging technology only enables postoperative outcome confirmation. In this study, we propose an approach to visualize short-burst low-pressure focal beam distribution that allows to be applied in FUS-BBB opening intervention on small animals. A backscattered acoustic-wave reconstruction method based on synchronization among focused ultrasound emission, diagnostic ultrasound receiving and passively beamformed processing were developed. We observed that focal beam could be successfully visualized for in vitro FUS exposure with 0.5-2 MHz without involvement of microbubbles. The detectable level of FUS exposure was 0.467 MPa in pressure and 0.05 ms in burst length. The signal intensity (SI) of the reconstructions was linearly correlated with the FUS exposure level both in-vitro (r(2) = 0.9878) and in-vivo (r(2) = 0.9943), and SI level of the reconstructed focal beam also correlated with the success and level of BBB-opening. The proposed approach provides a feasible way to perform real-time and closed-loop control of FUS-based brain drug delivery.

Introduction

Purpose Drug delivery with BBB opening
Study Objective Develop and validate a real-time backscattered acoustic-wave reconstruction method to visualize short-burst, low-pressure focused ultrasound focal beam distribution for guiding FUS-induced blood–brain barrier opening in small animals.
Animal model / Human subject Rat, adult, male
MRI or image guidance method Diagnostic ultrasound (passive beamforming/backscattered acoustic-wave reconstruction to visualize focal beam)
Targeted brain region(s) Hemisphere

Outcomes and Safety

Summary of Outcomes Synchronized backscattered acoustic-wave reconstruction enabled real-time visualization of burst mode, low pressure FUS exposure and guided successful ultrasound-induced BBB opening in rats, with reconstruction signal intensity linearly correlating with exposure level. BBB opening was observed at 0.467 MPa, while higher pressure produced pronounced opening accompanied by erythrocyte extravasation
Safety-related matter Erythrocyte extravasation occurred at 0.705 Mpa but not 0.467 Mpa, indicating hemorrhagic effects only at the higher exposure level.

Brain Region

Ultrasound Parameters

Ultrasound instrument Center-opened 1.5 MHz FUS transducer
FUS Frequency 1.5 MHz
FUS Pressure 0.11-0.705 Mpa
FUS Mode pulsed
Pulse duration 10 ms
Duration of a single FUS session 60 s
Focal Characteristics focal depth: None; focal length: None; aperture size: None
Treatment frequency Single

We are open to feedback. If you see a mistake or have a suggestion, please contact us.

← Back to Search