Opening the Blood-Brain Barrier with MR Imaging-guided Focused Ultrasound: Preclinical Testing on a Trans-Human Skull Porcine Model.
Authors: Huang Y, Alkins R, Schwartz ML, Hynynen K
Purpose To develop and test a protocol in preparation for a clinical trial on opening the blood-brain barrier (BBB) with magnetic resonance (MR) imaging-guided focused ultrasound for the delivery of chemotherapy drugs to brain tumors. Materials and Methods The procedures were approved by the institutional animal care committee. A trans-human skull porcine model was designed for the preclinical testing. Wide craniotomies were applied in 11 pigs (weight, approximately 15 kg). A partial human skull was positioned over the animal's brain. A modified clinical MR imaging-guided focused ultrasound brain system was used with a 3.0-T MR unit. The ultrasound beam was steered during sonications over a 3 × 3 grid at 3-mm spacing. Acoustic power levels of 3-20 W were tested. Bolus injections of microbubbles at 4 μL/kg were tested for each sonication. Levels of BBB opening, hemorrhage, and cavitation signal were measured with MR imaging, histologic examination, and cavitation receivers, respectively. A cavitation safety algorithm was developed on the basis of logistic regression of the measurements and tested to minimize the risk of hemorrhage. Results BBB openings of approximately 1 cm<sup>3</sup> in volume were visualized with gadolinium-enhanced MR imaging after sonication at an acoustic power of approximately 5 W. Gross examination of histologic specimens helped confirm Evans blue (bound to macromolecule albumin) extravasation, and hematoxylin-eosin staining helped detect only scattered extravasation of red blood cells. In cases where cavitation signals were higher than thresholds, sonications were terminated immediately without causing hemorrhage. Conclusion With a trans-human skull porcine model, this study demonstrated BBB opening with a 230-kHz system in preparation for a clinical trial. <sup>©</sup> RSNA, 2016 Online supplemental material is available for this article.
Introduction
Purpose
Drug delivery with BBB opening
Study Objective
To develop and test a protocol for MR-guided focused ultrasound–mediated blood–brain barrier opening in preparation for a clinical trial of chemotherapy delivery to brain tumors.
Animal model / Human subject
Porcine (pig, Sus scrofa domesticus); strain: not reported; age: not reported (weight ≈15 kg); sex: not reported
Disease model
healthy
MRI or image guidance method
MR imaging–guided focused ultrasound using a modified clinical system with a 3.0‑T MR unit; the ultrasound beam was steered during sonications over a 3×3 grid with 3 mm spacing and BBB opening was confirmed with gadolinium‑enhanced MR imaging
Targeted brain region(s)
Two Layers Of Depth Were
Targeted: Two To Four Spots At The Layer
Superior To The Ventricle In The Cortex
And Two To Four Spots Inferior To The Ven-
Tricle.
Cargo name and characteristics
230-kHz MR-guided focused ultrasound (physical/experimental modality used to open the BBB); gadolinium-based MRI contrast agent (small-molecule GBCA used to visualize BBB opening); Evans blue dye bound to serum albumin (small-molecule dye complexed with a protein macromolecule used to confirm extravasation)
Route of administration
Intravenous (bolus injections of microbubbles)
Outcomes and Safety
Summary of Outcomes
In a trans–human skull porcine model, MR-guided 230-kHz focused ultrasound produced ~1 cm³ blood–brain barrier openings (shown by gadolinium enhancement and Evans blue/albumin extravasation) at ~5 W with only scattered red blood cell extravasation and no hemorrhage when cavitation safety thresholds were employed.
Safety-related matter
Histology showed only scattered red blood cell extravasation (no hemorrhage), and sonications with high cavitation signals were immediately terminated by a cavitation safety algorithm to minimize hemorrhage risk.
Brain Region
Ultrasound Parameters
Ultrasound instrument
ExAblate 4000
FUS Frequency
0.23 MHz
FUS Intensity
NR
FUS Pressure
0.5 MPa
FUS Mode
pulsed
Pulse duration
2 ms
Duration of a single FUS session
50 sec
Focal Characteristics
Natural focus of array at 230 kHz: 3 mm diameter
Treatment frequency
multiple sessions