Design of patient-specific focused ultrasound arrays for non-invasive brain therapy with increased trans-skull transmission and steering range.

Authors: Hughes A, Hynynen K

The use of a phased array of ultrasound transducer elements to sonicate through the skull has opened the way for new treatments and the delivery of therapeutics beyond the blood-brain barrier. The limited steering range of current clinical devices, particularly at higher frequencies, limits the regions of the brain that are considered treatable by ultrasound. A new array design is introduced that allows for high levels of beam steering and increased transmission throughout the brain. These improvements are achieved using concave transducers normal to the outer-skull surface in a patient-specific configuration to target within the skull, so that the far-field of each beam is within the brain. It is shown that by using pulsed ultrasound waves timed to arrive in-phase at the desired target, sufficient levels of acoustic energy are delivered for blood-brain barrier opening throughout the brain.

Introduction

Purpose Transcranial ultrasound stimulation
Study Objective To design and evaluate a patient-specific transcranial FUS array to increase trans-skull transmission and extend the brain steering range.
Animal model / Human subject human cadaver skull
MRI or image guidance method CT
Targeted brain region(s) Lateral Brain Regions
Target coordinates AP: 60 mm; LR: 0 mm; IS: 0 mm

Outcomes and Safety

Summary of Outcomes The conformal patient-specific array improved trans-skull transmission, steering range, and focal quality compared with hemispherical arrays. Stimulations demonstrated acoustic pressures sufficient for potential BBB opening and drug delivery in the brain.
Safety-related matter Skull heating negligible; off target energy minimized at duty cycle less than 10% with short pulses.

Brain Region

Ultrasound Parameters

Ultrasound instrument Patient-specific concave focused ultrasound phased array transducer
FUS Frequency 500 kHz
FUS Pressure 0.24-1.4 Mpa
FUS Mode pulsed
Pulse duration 6 us, 10 us, 20 us
Duration of a single FUS session 2 ms
Focal Characteristics Focal depth: None; Focal length: 10 mm; Aperture size: 20 mm