Design of patient-specific focused ultrasound arrays for non-invasive brain therapy with increased trans-skull transmission and steering range.
Authors: Hughes A, Hynynen K
The use of a phased array of ultrasound transducer elements to sonicate through the skull has opened the way for new treatments and the delivery of therapeutics beyond the blood-brain barrier. The limited steering range of current clinical devices, particularly at higher frequencies, limits the regions of the brain that are considered treatable by ultrasound. A new array design is introduced that allows for high levels of beam steering and increased transmission throughout the brain. These improvements are achieved using concave transducers normal to the outer-skull surface in a patient-specific configuration to target within the skull, so that the far-field of each beam is within the brain. It is shown that by using pulsed ultrasound waves timed to arrive in-phase at the desired target, sufficient levels of acoustic energy are delivered for blood-brain barrier opening throughout the brain.
Introduction
Purpose
Transcranial ultrasound stimulation
Study Objective
To design and evaluate a patient-specific transcranial FUS array to increase trans-skull transmission and extend the brain steering range.
Animal model / Human subject
human cadaver skull
MRI or image guidance method
CT
Targeted brain region(s)
Lateral Brain Regions
Target coordinates
AP: 60 mm; LR: 0 mm; IS: 0 mm
Outcomes and Safety
Summary of Outcomes
The conformal patient-specific array improved trans-skull transmission, steering range, and focal quality compared with hemispherical arrays. Stimulations demonstrated acoustic pressures sufficient for potential BBB opening and drug delivery in the brain.
Safety-related matter
Skull heating negligible; off target energy minimized at duty cycle less than 10% with short pulses.
Brain Region
Ultrasound Parameters
Ultrasound instrument
Patient-specific concave focused ultrasound phased array transducer
FUS Frequency
500 kHz
FUS Pressure
0.24-1.4 Mpa
FUS Mode
pulsed
Pulse duration
6 us, 10 us, 20 us
Duration of a single FUS session
2 ms
Focal Characteristics
Focal depth: None; Focal length: 10 mm; Aperture size: 20 mm