Influence of Nanobubble Concentration on Blood-Brain Barrier Opening Using Focused Ultrasound Under Real-Time Acoustic Feedback Control.
Authors: Cheng B, Bing C, Xi Y, Shah B, Exner AA, Chopra R
Real-time acoustic feedback control based on harmonic emissions of stimulated microbubbles may serve as a way to achieve reliable blood-brain barrier (BBB) opening with focused ultrasound in the brain. Previously, we demonstrated BBB opening was possible using sub-micron bubbles (aka nanobubbles) and produced comparable results to commercially available microbubbles (Optison, Definity, etc.). The harmonic emissions and acoustic control were observed to be more consistent using nanobubbles, which warrants further study of BBB opening using these agents. This study examined the stimulated acoustic emissions of nanobubbles at different concentrations both in vitro and in vivo and evaluated BBB opening under real-time acoustic feedback control across concentrations. Original nanobubbles (10<sup>11</sup> bubbles/mL) have long in vitro persistence (7.3 ± 3.3 min) and circulation time in rats (approximately 10 min) under exposures in this study, and both degraded with dilutions. With all three tested dilutions (1:1, 1:10 and 1:100), successful BBB opening was reliably achieved under real-time feedback control.
Introduction
Purpose
Drug delivery with BBB opening
Study Objective
To investigate how varying nanobubble concentration influences focused ultrasound-mediated blood–brain barrier opening under real-time acoustic feedback control.
Animal model / Human subject
Sprague-Dawley rats (sex: female)
Disease model
Healthy
Cargo name and characteristics
Nanobubbles, Evans blue
Route of administration
intravenous
Outcomes and Safety
Summary of Outcomes
All tested dilutions of nanobubbles (1:1, 1:10, and 1:100) were able to open the BBB reliably.
Safety-related matter
A safety evaluation was not included in the study.
Brain Region
Ultrasound Parameters
Ultrasound instrument
custom single-element focused lead zirconate titanate (PZT) transducer (DL-54, DeL Piezo Specialties, LLC, West Palm Beach, FL, USA)
FUS Frequency
0.5 MHz
FUS Pressure
0.14-0.76
FUS Mode
pulsed
Pulse duration
10 ms
Duration of a single FUS session
120 s
Focal Characteristics
diameter: 75 mm, radius of curvature: 60 mm
Treatment frequency
single
We are open to feedback. If you see a mistake or have a suggestion, please contact us.
← Back to Search