A Magnetic Resonance-Guided Focused Ultrasound Neuromodulation System With a Whole Brain Coil Array for Nonhuman Primates at 3 T.
Authors: Li Y, Lee J, Long X, Qiao Y, Ma T, He Q, Cao P, Zhang X, Zheng H
The phased-array radio frequency (RF) coil plays a vital role in magnetic resonance-guided focused ultrasound (MRgFUS) neuromodulation studies, where accurate brain functional stimulations and neural circuit observations are required. Although various designs of phased-array coils have been reported, few are suitable for ultrasound stimulations. In this study, an MRgFUS neuromodulation system comprised of a whole brain coverage non-human primate (NHP) RF coil and an MRI-compatible ultrasound device was developed. When compared to a single loop coil, the NHP coil provided up to a 50% increase in the signal-to-noise ratio (SNR) in the brain and acquired better anatomical image-quality. The NHP coil also demonstrated the ability to achieve higher spatial resolution and reduce distortion in echo-planer imaging (EPI). Ultrasound beam characteristics and transcranial magnetic resonance acoustic radiation force (MR-ARF) were measured for simulated positions, and calculated B<sub>0</sub> maps were employed to establish MRI-compatibility. The differences between focused off and on ultrasound techniques were measured using SNR, g-factors, and temporal SNR (tSNR) analyses and all deviations were under 2.3%. The EPI images quality and stable tSNR demonstrated the suitability of the MRgFUS neuromodulation system to conduct functional MRI studies. Last, the time course of the blood oxygen level dependent (BOLD) signal of posterior cingulate cortex in a focused ultrasound neuromodulation study was detected and repeated with MR thermometry.
Introduction
Purpose
Transcranial ultrasound stimulation
Study Objective
To develop and evaluate an MR-guided focused ultrasound neuromodulation system with a whole-brain coil array for nonhuman primates at 3 T.
Animal model / Human subject
Non-human primates: Rhesus monkey (Macaca mulatta, age: 7 years, sex: male)
Disease model
Healthy
MRI or image guidance method
3T MRI system (uMR 790, Shanghai United Imaging Healthcare, Shanghai, China)
Targeted brain region(s)
Posterior Cingulate Cortex
Outcomes and Safety
Summary of Outcomes
The NHP coil had an improved performance to a single coil, increasing the signl--to-noise ration and achieving better image quality. The MRgFUS system (with the NHP phased array RF coil) was demonstrated as viable for neuromodulation.
Brain Region
Ultrasound Parameters
Ultrasound instrument
custom-designed, homemade MRI-compatible ultrasound transducer
FUS Frequency
0.3 MHz
FUS Pressure
0.758 MPa
FUS Mode
pulsed
Pulse duration
0.5 ms
Duration of a single FUS session
1 s
Focal Characteristics
aperature: 60 mm
Treatment frequency
single