Suppressing Grating Lobes for Transcranial Focused Ultrasound System by Frequency-Modulated Excitation.
Authors: Qu X, Shen G, Wu N, Wu H, Qiao S, Wang E, Chen Y, Wang H
Transcranial focused ultrasound is a novel noninvasive therapeutic modality for glioblastoma and other disorders of the brain. However, because the phase aberrations caused by the skull need to be corrected with computed tomography (CT) images, the transcranial transducer is tightly fixed on the patient's head to avoid any variation in the relative position, and the focus shifting relies mainly on the capacity for electronic beam steering. Due to the presence of grating lobes and the rapid degradation of the focus quality with increasing focus-shifting distance, transcranial focus-shifting sonication may damage healthy brain tissue unintentionally. To reduce the risks associated with transcranial focused ultrasound therapy, linear frequency-modulated (FM) excitation is proposed. The k-space corrected pseudospectral time domain (PSTD) and acoustic holography approach based on the Rayleigh integral are combined to calculate the distribution of the deposited acoustic power. The corresponding simulation was performed with axial/lateral focus shifting at different distances. The distributions of the deposited acoustic power show that linear FM excitation can effectively suppress undesired prefocal grating lobes without compromising focus quality.
Introduction
Purpose
Transcranial ultrasound stimulation
Study Objective
To reduce grating lobes in transcranial focused ultrasound systems by applying frequency-modulated excitation.
Outcomes and Safety
Summary of Outcomes
No biological or behavioral effects were reported; the study demonstrated that frequency-modulated excitation schemes successfully suppressed grating lobes in a transcranial focused ultrasound system.
Safety-related matter
No safety-related information or adverse effects are mentioned in the provided text; only the paper title was provided.
Brain Region
Ultrasound Parameters
Focal Characteristics
Focal depth: None; Focal length: None; Aperture size: None
We are open to feedback. If you see a mistake or have a suggestion, please contact us.
← Back to Search