Blood-brain barrier opening with focused ultrasound in Parkinson's disease dementia.

Authors: Gasca-Salas C, Fernández-Rodríguez B, Pineda-Pardo JA, Rodríguez-Rojas R, Obeso I, Hernández-Fernández F, Del Álamo M, Mata D, Guida P, Ordás-Bandera C, Montero-Roblas JI, Martínez-Fernández R, Foffani G, Rachmilevitch I, Obeso JA

MR-guided focused ultrasound (MRgFUS), in combination with intravenous microbubble administration, has been applied for focal temporary BBB opening in patients with neurodegenerative disorders and brain tumors. MRgFUS could become a therapeutic tool for drug delivery of putative neurorestorative therapies. Treatment for Parkinson's disease with dementia (PDD) is an important unmet need. We initiated a prospective, single-arm, non-randomized, proof-of-concept, safety and feasibility phase I clinical trial (NCT03608553), which is still in progress. The primary outcomes of the study were to demonstrate the safety, feasibility and reversibility of BBB disruption in PDD, targeting the right parieto-occipito-temporal cortex where cortical pathology is foremost in this clinical state. Changes in β-amyloid burden, brain metabolism after treatments and neuropsychological assessments, were analyzed as exploratory measurements. Five patients were recruited from October 2018 until May 2019, and received two treatment sessions separated by 2-3 weeks. The results are set out in a descriptive manner. Overall, this procedure was feasible and reversible with no serious clinical or radiological side effects. We report BBB opening in the parieto-occipito-temporal junction in 8/10 treatments in 5 patients as demonstrated by gadolinium enhancement. In all cases the procedures were uneventful and no side effects were encountered associated with BBB opening. From pre- to post-treatment, mild cognitive improvement was observed, and no major changes were detected in amyloid or fluorodeoxyglucose PET. MRgFUS-BBB opening in PDD is thus safe, reversible, and can be performed repeatedly. This study provides encouragement for the concept of BBB opening for drug delivery to treat dementia in PD and other neurodegenerative disorders.

Introduction

Purpose Drug delivery with BBB opening
Study Objective To assess the safety, feasibility, and reversibility of MR-guided focused ultrasound-induced blood–brain barrier opening in the right parieto-occipito-temporal cortex of patients with Parkinson's disease dementia.
Animal model / Human subject Homo sapiens (human patients with Parkinson's disease with dementia); strain: N/A; age: 60–80 years; sex: None
Disease model Parkinson's disease with dementia (PDD)
MRI or image guidance method MR-guided (ExAblate Neuro) with a stereotactic frame; intraprocedural MRI (T1, T2, T2*/GRE) and MR thermometry used for planning, target selection and real-time monitoring.
Targeted brain region(s) Parieto-Occipito-Temporal Cortex
Route of administration intravenous

Outcomes and Safety

Summary of Outcomes In a phase I trial in five PDD patients, MR‑guided focused ultrasound with intravenous microbubbles produced safe, reversible BBB opening in 8/10 treatments with no serious adverse events, transient minor side effects and heterogeneous mild cognitive improvements, using two sessions (2–3 weeks apart) with mean sonications and maximum sonication power of 5.2 sonications (max 17.4 W) for stage 1 and 5.8 sonications (max 19.4 W) for stage 2.
Duration of biological effect 24 h to 1 week
Safety-related matter Overall, MRgFUS BBB opening was feasible, reversible, and safe with no serious clinical or radiological adverse events (no hemorrhage or edema) and no worsening of PD; all patients were discharged the morning after and follow-up (7–12 months) showed no new adverse events. Minor transient adverse effects included local phlebitis/needle-site redness, anxiety and transient blood-pressure elevation, brief restlessness and transient confusion likely related to sedation, and transient SWAN hypointensities in three patients (one persisting/attenuated up to 2 months) that may reflect small blood extravasation but had no clinical correlate.

Brain Region

Ultrasound Parameters

Ultrasound instrument ExAblate Neuro (InSightec, Haifa); 1024-element phased-array transducer, 220 kHz center frequency; transducer aperture/diameter: None
FUS Frequency 220 kHz
FUS Mode pulsed
Focal Characteristics focal depth: None; focal length: None; aperture size: None
Treatment frequency multiple sessions