Comparison Between Ray-Tracing and Full-Wave Simulation for Transcranial Ultrasound Focusing on a Clinical System Using the Transfer Matrix Formalism.

Authors: Bancel T, Houdouin A, Annic P, Rachmilevitch I, Shapira Y, Tanter M, Aubry JF

Only one high-intensity focused ultrasound device has been clinically approved for transcranial brain surgery at the time of writing. The device operates within 650 and 720 kHz and corrects the phase distortions induced by the skull of each patient using a multielement phased array. Phase correction is estimated adaptively using a proprietary algorithm based on computed-tomography (CT) images of the patient's skull. In this article, we assess the performance of the phase correction computed by the clinical device and compare it to: 1) the correction obtained with a previously validated full-wave simulation algorithm using an open-source pseudo-spectral toolbox and 2) a hydrophone-based correction performed invasively to measure the aberrations induced by the skull at 650 kHz. For the full-wave simulation, three different mappings between CT Hounsfield units and the longitudinal speed of sound inside the skull were tested. All methods are compared with the exact same setup due to transfer matrices acquired with the clinical system for N = 5 skulls and T = 2 different targets for each skull. We show that the clinical ray-tracing software and the full-wave simulation restore, respectively, 84% ± 5% and 86% ± 5% of the pressure obtained with hydrophone-based correction for targets located in central brain regions. On the second target (off-center), we also report that the performance of both algorithms degrades when the average incident angles of the acoustic beam at the skull surface increase. When incident angles are higher than 20°, the restored pressure drops below 75% of the pressure restored with hydrophone-based correction.

Introduction

Purpose Transcranial ultrasound stimulation
Study Objective To compare the accuracy and performance of ray-tracing versus full-wave simulations for transcranial ultrasound focusing on a clinical system using the transfer matrix formalism.

Outcomes and Safety

Summary of Outcomes No biological or behavioral effects were reported; the paper is a methods comparison of ray-tracing versus full-wave simulation for transcranial ultrasound focusing using the transfer matrix formalism and does not report any specific focused-ultrasound parameter sets as being successful.
Safety-related matter No safety-related statements or adverse effects are mentioned in the provided text; only the paper title was included.

Brain Region

Visualization unavailable

Ultrasound Parameters

Focal Characteristics Focal depth: None; Focal length: None; Aperture size: None