Large-Volume Focused-Ultrasound Mild Hyperthermia for Improving Blood-Brain Tumor Barrier Permeability Application.

Authors: Chan H, Chang HY, Lin WL, Chen GS

Mild hyperthermia can locally enhance permeability of the blood-tumor barrier in brain tumors, improving delivery of antitumor nanodrugs. However, a clinical transcranial focused ultrasound (FUS) system does not provide this modality yet. The study aimed at the development of the transcranial FUS technique dedicated for large-volume mild hyperthermia in the brain. Acoustic pressure, multiple-foci, temperature and thermal dose induced by FUS were simulated in the brain through the skull. A 1-MHz, 114-element, spherical helmet transducer was fabricated to verify large-volume hyperthermia in the phantom. The simulated results showed that two foci were simultaneously formed at (2, 0, 0) and (-2, 0, 0) and at (0, 2, 0) and (0, -2, 0), using the phases of focusing pattern 1 and the phases of focusing pattern 2, respectively. Switching two focusing patterns at 5 Hz produced a hyperthermic zone with an ellipsoid of 7 mm × 6 mm × 11 mm in the brain and the temperature was 41-45 °C in the ellipsoid as the maximum intensity was 150 W/cm<sup>2</sup> and sonication time was 3 min. The phased array driven by switching two mode phases generated a 41 °C-contour region of 10 ± 1 mm × 8 ± 2 mm × 13 ± 2 mm in the phantom after 3-min sonication. Therefore, we have demonstrated our developed FUS technique for large-volume mild hyperthermia.

Introduction

Purpose Drug delivery WITHOUT BBB opening
Study Objective To develop a transcranial focused ultrasound (FUS) technique for producing large-volume mild hyperthermia in the brain.
Disease model brain tumors
MRI or image guidance method None (targeting achieved by electronic phased-array focusing using precomputed phase patterns/simulations with a 114-element spherical helmet transducer; steering via phase patterns and mode switching)
Target coordinates (2, 0, 0), (-2, 0, 0), (0, 2, 0), (0, -2, 0)

Outcomes and Safety

Summary of Outcomes Transcranial phased-array FUS produced localized mild hyperthermia (41–45 °C) in a roughly ellipsoidal brain region (simulated 7×6×11 mm; phantom ~10±1×8±2×13±2 mm) after 3 min sonication, supporting potential to enhance blood‑tumor barrier permeability. Successful FUS parameters were: 1 MHz, 114‑element spherical helmet transducer, switching two phase-focused patterns (foci at (±2,0,0) and (0,±2,0)) at 5 Hz, peak intensity ~150 W/cm², 3 min sonication.
Duration of biological effect 3 min
Safety-related matter No adverse effects or safety issues were reported; the study was limited to simulations and phantom experiments demonstrating targeted mild hyperthermia (41–45 °C) in small ellipsoid volumes after 3-minute sonication.

Brain Region

Visualization unavailable

Ultrasound Parameters

Ultrasound instrument 1-MHz, 114-element spherical helmet transducer; Manufacturer: None; Aperture/diameter: None
FUS Frequency 1 MHz
FUS Intensity 150 W/cm2
FUS Mode pulsed
Duration of a single FUS session 3 min
Focal Characteristics Focal depth: None; Focal length: None; Aperture size: None
Treatment frequency single

We are open to feedback. If you see a mistake or have a suggestion, please contact us.

← Back to Search