Pitt Shield

Factors Associated with Heating Efficiency in Transcranial Focused Ultrasound Therapy.

Authors: Yamamoto K, Ito H, Fukutake S, Odo T, Kamei T, Yamaguchi T, Taira T

Transcranial magnetic resonance-guided focused ultrasound (FUS) therapy is a less invasive stereotactic treatment for tremor and other movement disorders. A sufficiently high temperature in the target brain tissue is crucial during ablation procedures for good outcomes. Therefore, maximizing the heating efficiency is critical in cases where high temperature cannot be achieved because of patient-related characteristics. However, a strategy to achieve the desired therapeutic temperature with FUS has not yet been established. This study aimed to investigate the procedural factors associated with heating efficiency in FUS.We retrospectively reviewed and analyzed data from patients who underwent FUS for ventralis intermedius (VIM) nucleus thalamotomy. In all, 30 consecutive patients were enrolled. 18 with essential tremor (ET), 11 with tremor-dominant Parkinson's disease (TDPD), and 1 with Holmes tremor. Multivariate regression analysis showed that decline in heating efficiency was associated with lower skull density ratio (SDR) and a greater subtotal rise in temperature until the previous sonication. To maximize heating efficiency, the temperature increase should be set to the least value in the target alignment and verification phases, and subsequently should be increased sufficiently in the treatment phase. This strategy may be particularly beneficial in cases where high ablation temperatures cannot be achieved because of patient-related characteristics. Importantly, a broad patient population would benefit from this strategy as it could reduce the need for high energy to achieve therapeutic temperatures, thereby decreasing the risks of adverse events.

Introduction

Purpose Other
Study Objective ventral intermediate nucleus of the thalamus
Animal model / Human subject Homo sapiens
Disease model essential tremor
MRI or image guidance method MRI-guided with stereotactic frame
Targeted brain region(s) Vim
Target coordinates 12 mm lateral to 3rd ventricle wall

Outcomes and Safety

Summary of Outcomes 650 kHz MRgFUS heating efficiency in ET patients is limited by low skull density ratio (SDR) and prior sub-therapeutic heating.
Duration of biological effect permanent
Safety-related matter Low SDR correlates with higher risks of skull heating and skin burns.

Brain Region

Ultrasound Parameters

Ultrasound instrument ExAblate Neuro 4000
FUS Frequency 650 kHz
FUS Intensity Not reported in the provided text
FUS Pressure Not reported
FUS Mode continuous
Pulse duration 20 s
Duration of a single FUS session 120 min
Focal Characteristics Not reported
Treatment frequency single session

We are open to feedback. If you see a mistake or have a suggestion, please contact us.